The components of uncertainty that can be estimated from the data are listed in Table A8, together with those that cannot. The estimates are relative standard uncertainties of a single α- or β-acid content and are given separately for the two bottles.
| Component | Bottle A | Bottle B | Type / direction | Basis of the estimate |
|---|---|---|---|---|
| Mass of hop sample | 0.05 | 0.05 | random | S1 masses recorded to 0.1 mg, S2 to 0.01 mg, ~100 mg samples; balance repeatability ~0.02 mg assumed. |
| Mass of DMS in the reference solution | 0.68 | 0.12 | random (per bottle, systematic for all samples in that bottle) | A: "4.5 mg" recorded to two significant figures (±0.05 mg rounding, rectangular) plus 0.01 mg balance; B: 8.74 mg. |
| Purity of DMS | 0.10 | 0.10 | systematic | Taken as 0.9998 from the PNA terpene-assay worksheet; certificate not part of this data set. |
| Solvent volume, sample tube vs reference tube (800 µL each) | 1.40 | 1.40 | random | Assumed 1 % repeatability per delivery of CDCl₃ with the same pipette; a common delivery bias cancels. |
| Relaxation (d1 = 15 s; pulse angle and acquisition time not recorded) | 1.00 | 1.00 | systematic, sign uncertain | DMS T₁ 4–5 s in CDCl₃ (BIPM ISRD-02); α-acid 3-OH T₁ about 4 s; TMS saturation cancels only if T₁(TMS) is the same in reference and sample. |
| Integration: TMS window 0.24 ppm (samples) vs 0.085/0.134 ppm (references) | 2.70 | 1.40 | systematic, analyte LOW | Lorentzian TMS lines (1.5/1.9 Hz) in the reference spectra are integrated over narrower windows; estimated from the digitized reference traces. |
| Integration: DMS ¹³C satellites outside the DMS window | 1.10 | 1.10 | systematic, analyte HIGH | 1.07 % of DMS intensity lies in satellites at δ 2.75 and 3.21 (outside the window); hydroxyl signals have no satellites. |
| Integration: truncation of analyte lines vs inclusion of minor lines | 1.50 | 1.50 | systematic, small | Line fitting of the digitized spectra: Σ fitted α lines / Σ α windows median 0.985 (IQR 0.947–0.994); for β 0.93, because the β windows also collect minor lines. |
| Loss of TMS (volatility) between reference and sample preparation | not quantified | not quantified | systematic, sign depends on history | Not quantifiable from the data. Bottle A contained 1.29 mM TMS against 2.85 mM in the new bottle B, consistent with loss of TMS from an opened bottle; loss from a sample tube during extraction biases results HIGH. |
| H/D exchange and exchange broadening of the enolic OH | not quantified | not quantified | systematic, analyte LOW | Hoek et al. found the α-acid 3-OH 9–12 % below the prenyl signals (possibly relaxation). ICE-4: NMR α/β 6.5 % below the certificate. |
| Completeness of a single 800 µL extraction | not quantified | not quantified | systematic, analyte LOW | Never verified by re-extraction. |
| Combined (quantified components only), relative standard uncertainty | 3.77 | 2.90 | Root-sum-square of the quantified random and systematic terms above; see text. |
The weighing of the hop sample contributes negligibly. The mass of DMS in the calibration solution of bottle A, recorded as 4.5 mg with two significant figures, contributes 0.7 %, almost entirely through its rounding, against 0.1 % for the 8.74 mg of bottle B, and a larger weighed mass recorded to 0.01 mg would remove the term. The purity of the DMS was taken as 0.9998, the value entered in the PNA terpene-assay worksheet; dimethyl sulfone is well established as a qNMR calibrant and is the subject of a BIPM internal-standard reference data sheet, which gives its longitudinal relaxation time in CDCl₃ as 4–5 s and its solubility as greater than 5 mg mL⁻¹.21–23 The bottle B solution, at 10.9 mg mL⁻¹, is above that figure, and the complete dissolution of the calibrant should be confirmed visually in any repetition. The volumes of the calibration and sample solutions enter as a ratio and were assumed to be equal; a repeatability of 1 % per delivery of a dense, volatile solvent by pipette gives 1.4 % for the ratio, while a common bias of the pipette cancels.
The relaxation conditions are only partly known. With a 15 s relaxation delay, DMS (T₁ 4–5 s) and the enolic protons (T₁ about 4 s for the α-acid 3-OH1) are fully relaxed at small pulse angles and relaxed to about 97–98 % at 90°, depending on the acquisition time, which was not recorded; TMS relaxes more slowly, but its saturation cancels provided that its T₁ is the same in the calibration and sample solutions, which is plausible but has not been checked. A standard uncertainty of 1 % has been assigned. Three integration effects were quantified from the digitized spectra. The TMS signal was integrated over δ 0.12 to −0.12 in the sample spectra but over narrower windows in the two calibration spectra (0.085 and 0.134 ppm), which, for the Lorentzian TMS lines of 1.5–1.9 Hz observed there, capture 1.4–2.7 % less of the line and bias the analyte results low by that amount. The DMS window excludes the ¹³C satellites of DMS at δ 2.75 and 3.21 (1.07 % of its intensity), whereas the hydroxyl signals have none, and it truncates the wings of the DMS line by a further 0.8 %, so that the DMS integral is 1.8 % low and the results correspondingly high. The truncation of the analyte lines by their windows and the inclusion of neighboring minor lines nearly cancel (Section A3.3); referred to the fitted principal lines, the α-acid window integrals are 1.5 % high. Treated as corrections, the three effects would change the α-acid contents by −0.7 % in bottle A and −2.0 % in bottle B, which is within the uncertainty; they have not been applied. The root-sum-square of the quantified components is 3.8 % for bottle A and 2.9 % for bottle B.
Three further components could not be quantified, and any one of them may exceed the combined figure. The first and most serious is the stability of the TMS content. The transfer standard is volatile, bottle A contains less than half the TMS of bottle B, most probably because TMS has been lost from the opened bottle, and neither the interval between the calibration and the extractions nor the time for which the extraction vials stood open, were sonicated or were filtered is recorded; TMS lost from a sample tube before acquisition biases that result high, and TMS lost from the bottle between extraction and calibration biases all results from that bottle low. The second is exchange of the enolic protons with the deuterium of the solvent and its impurities, which can only lower the hydroxyl integrals and which the ICE-4 α/β ratio bounds at about 7 % for the α-acids relative to the β-acids. The third is the completeness of a single extraction of 100 mg of hops into 800 µL of CDCl₃, which has never been tested by re-extraction and can only lower the results.
The empirical evidence is consistent with a combined uncertainty of the order of 5–10 %. The thirteen S2 samples agree with their declared α-acid contents with a root-mean-square difference of 1.84 % α-acids and a standard deviation of the ratio of 12.5 %, but that spread includes the uncertainty of the declared values, whose method and date are unknown, and cannot be apportioned. The failure of the ICE-4 check (Section A4.5) removes the one independent test of the absolute scale, and until it has been repeated, the absolute contents reported here should be regarded as provisional to within about ±10 %, the relative quantities, homolog fractions and α/β ratios as well founded, and the comparison of samples measured with the same bottle as more reliable than a comparison across bottles.