Process NMR Associates
PNA-WP-2026-HOPS-01-A1 · Section A4

A4. Direct qNMR through a TMS-Transferred Dimethyl Sulfone Calibration

NMR Spectroscopy in the Analysis of Hops (Humulus lupulus L.) · Addendum A1: Direct qNMR and Chemometrics

A4.1 Principle

In the classical internal-standard form of quantitative ¹H NMR, a weighed calibrant is dissolved in the same solution as the analyte, and the content follows from the ratio of the two integrals, the numbers of protons, the molar masses and the masses weighed, without calibration against the analyte itself.10,11 The procedure used here separates the calibrant from the sample. The TMS contained in the deuterated solvent is present in both the calibration solution and the sample solution, and if the two are taken from the same bottle in the same volume, the TMS signal of each spectrum carries the same number of protons. Setting it to ten units in every spectrum places all the spectra on one scale, and a single calibration spectrum per bottle, in which a weighed quantity of DMS is dissolved in 800 µL of that bottle, converts the scale into moles. The approach belongs to the family of methods in which a signal of the solvent, rather than a weighed addition, serves as the concentration reference, whether it is the signal of a protonated solvent such as water12 or the residual signal of a deuterated solvent,13 and it shares with external calibration methods, of which PULCON and ERETIC are the best known,14,15 the requirement that the link between calibration spectrum and sample spectrum be established rather than assumed. Its sequence is summarized in Figure A4.

The TMS-transferred DMS calibration used for the direct qNMR of the 49 hop extracts
Figure A4. The TMS-transferred DMS calibration used for the direct qNMR of the 49 hop extracts. The TMS signal of each spectrum is set to ten units; the calibration factor of each bottle of CDCl₃ is obtained once from a solution of weighed DMS in 800 µL of that bottle.

For each bottle s, the calibration factor kₛ, the number of millimoles of ¹H represented by one normalized integral unit, is obtained from the DMS spectrum:

kₛ = N(DMS) · m(DMS)ₛ · P(DMS) / [M(DMS) · I(DMS)ₛ] (A1)

in which N(DMS) = 6 is the number of DMS protons, m(DMS)ₛ the mass of DMS dissolved in 800 µL of the bottle, P(DMS) its purity (0.9998), M(DMS) = 94.128 g mol⁻¹ and I(DMS)ₛ its integral with TMS equal to ten. For an analyte x whose signal of Nₓ protons has the normalized integral Iₓ in the spectrum of a hop extract of mass m, the amount in the tube and the content of the hop follow as

nₓ = Iₓ · kₛ / Nₓ ; w(x) = nₓ · M(x) / m (A2)

which, written out in full, is the familiar internal-standard expression multiplied by the ratio of the solvent volumes in the sample and calibration tubes, a ratio taken as unity:

w(x) = [Iₓ / I(DMS)ₛ] · [N(DMS) / Nₓ] · [M(x) / M(DMS)] · [m(DMS)ₛ / m] · P(DMS) · [V(sample) / V(DMS)] (A3)

Every hydroxyl signal is a single proton (Nₓ = 1), as are the diene H-2 of myrcene and the H-3 of α-humulene. The α-acid homologs were calculated from W7, W8 and W9 with the molar masses of adhumulone and humulone (362.466 g mol⁻¹) and cohumulone (348.439), and the β-acid homologs from the sums of the major and minor tautomer windows, W1 + W13 (adlupulone, 414.586), W2 + W16 (colupulone, 400.559) and W3 + W14 (lupulone, 414.586), following Section A3.5. Myrcene-type diene was calculated as myrcene (136.238) and α-humulene with 204.357 g mol⁻¹. All contents are on the as-is basis, since no moisture determinations were made; the companion workbook performs every calculation with live formulas, so that the calibration constants can be edited.

A4.2 The Calibration of the Two Bottles

The two calibration spectra and the constants derived from them are given in Table A4. For the original bottle (A), 4.5 mg of DMS in 800 µL gave I(DMS) = 230.707, whence k_A = 1.2431 × 10⁻³ mmol per unit; for the new bottle (B), 8.74 mg gave I(DMS) = 203.484 and k_B = 2.7373 × 10⁻³ mmol per unit. Because k = 12 n(TMS)/10, the same numbers give the TMS content of each bottle: 1.04 µmol in 800 µL (1.29 mM) for bottle A and 2.28 µmol (2.85 mM) for bottle B, a ratio of 2.20.

Table A4. Calibration of the two bottles of CDCl₃ (TMS) against dimethyl sulfone (800 µL; TMS integral set to 10): calibration factor k (Eq. A1) and the TMS content implied by it.
BottleDMS (mg)I(DMS)Windows: DMS; TMS (ppm)k (mmol ¹H per unit)TMS in 800 µL (µmol)TMS (mM)Samples
A (original bottle)4.5230.7073.12–2.84; 0.03 to −0.051.2431 × 10⁻³1.0361.2921 (S1)
B (new CIL bottle)8.74203.4843.15–2.80; 0.05 to −0.082.7373 × 10⁻³2.2812.8515 (S1) + 13 (S2)
k = 6·m(DMS)·P/[M(DMS)·I(DMS)], P = 0.9998, M(DMS) = 94.128 g mol⁻¹; n(TMS) = 10·k/12. Ratio k(B)/k(A) = 2.202. Nominal TMS contents of commercial CDCl₃: 0.03 % v/v ≈ 2.2 mM; 0.05 % v/v ≈ 3.7 mM.

The difference is the first important result of the addendum, because it shows that the TMS content of a bottle of CDCl₃ cannot be taken from its label. Commercial CDCl₃ is most commonly supplied with 0.03 or 0.05 % (v/v) of TMS, about 2.2 and 3.7 mM, and neither bottle contains a nominal amount. The TMS content of the older bottle is some 40 % below the lower of the two nominal grades and less than half that of the new bottle, a deficit that is most simply explained by the loss of TMS, which boils at about 27 °C, from a bottle that had been opened many times. The calibration spectrum of bottle A also shows a water signal at δ 1.59 far larger than that of bottle B, consistent with long use.16 The practical corollary is that the calibration of a bottle is valid only for as long as its TMS content is stable, and that the TMS content of an opened bottle must be expected to fall with time. The date of the bottle A calibration relative to the extractions made with it is not recorded, and this is the largest unquantified uncertainty of the method (Section A5).

A4.3 Assignment of the Samples to the Bottles and the Test of the 8.74 mg Assumption

A wrong assignment of a sample to a bottle changes its results by the factor 2.20 and is therefore easy to detect by comparison with the declared α-acid content, even though the declared values are nominal. The test is summarized in Table A5 and in Figure A5.

Table A5. Test of the assignment of the samples to the two solvent bottles: ratio of the α-acid content by direct qNMR to the declared α-acid content under the assigned bottle and under the alternative.
GroupnBottleMeanMedianSDRangeAlt. bottleMean (range) under alternative
S1 assigned to bottle A (incl. 022)21A0.7120.6770.2070.315–1.433B1.567 (0.693–3.156)
S1 assigned to bottle B15B0.7970.7620.1670.484–1.201A0.362 (0.220–0.545)
S2 pellets (043–050)8B1.0141.0370.1270.825–1.164A0.460 (0.375–0.529)
S2 Cryo pellets (038–042)5B1.0711.0140.1270.937–1.226A0.486 (0.426–0.557)
All S2 (038–050)13B1.0361.0300.1250.825–1.226A0.470 (0.375–0.557)
Sample 022 (duplicated 032 entry)1A0.830B1.827
S2 under bottle B: r = 0.975 between measured and declared α-acids, slope 1.14, intercept −1.08 %, RMS difference 1.84 % α-acids. S1 bottle A vs bottle B: Welch t-test p = 0.18. S1 vs S2: Mann–Whitney p = 1 × 10⁻⁵.

For samples 038–050, the assumption that bottle B was used (8.74 mg of DMS) gives α-acid contents whose ratio to the declared values has a mean of 1.036, a median of 1.030, a standard deviation of 0.125 and a range of 0.825–1.226 over the thirteen samples. Under the alternative assumption of bottle A, the ratio would be 0.470 (0.375–0.557), that is, every one of these recently purchased samples, including Cryo pellets sold on a certified analysis, would contain less than 56 % of its declared α-acids. The 8.74 mg assumption is therefore confirmed, the samples were retained in the quantitative analysis, and the instruction to withdraw them did not have to be applied. Across the thirteen samples the measured and declared α-acid contents are correlated with r = 0.975, a slope of 1.14 and an intercept of −1.08 % α-acids, and they differ by +0.76 % on average (standard deviation 1.74, root-mean-square difference 1.84 % α-acids).

(a) α-Acid contents by direct qNMR against the α-acid contents declared on the packages, by sample set and solvent bottle; the solid line is the line of equality and the dotted line marks one half of…
Figure A5. (a) α-Acid contents by direct qNMR against the α-acid contents declared on the packages, by sample set and solvent bottle; the solid line is the line of equality and the dotted line marks one half of the declared value. (b) Ratio of measured to declared α-acids for samples 038–050 under the two calibrations: bottle B (8.74 mg of DMS, adopted) and bottle A (4.5 mg).

For the S1 samples the assignments of the bottle workbook were tested in the same way. The 21 samples assigned to bottle A, including sample 022, whose entry had been duplicated as 032, gave a mean ratio of 0.712 (median 0.677; range 0.315–1.433), and the 15 assigned to bottle B a mean of 0.797 (median 0.762; range 0.484–1.201); the two groups do not differ significantly (Welch t-test, p = 0.18). Exchanging the bottles would raise the first group to 1.57 and lower the second to 0.36, and for sample 022 alone the ratio would become 1.83 against 0.83 as assigned. Every assignment of the workbook is thus consistent with the data, and 022 belongs to bottle A. The S1 samples as a whole, however, contain markedly less α-acid relative to their labels than the S2 samples (Mann–Whitney test, p = 1 × 10⁻⁵), about three quarters of the declared value on average.

That deficit is not a calibration artifact. It is found in both bottles; it is accompanied by an even larger deficit of hop oil, the median myrcene-type content being 0.31 % in S1 against 0.94 % in the S2 pellets; and losses of α-acids of this order during the storage of hops and pellets under non-ideal conditions are well documented, the loss being indexed by the hop storage index.17,18 A declared α-acid content is, moreover, a nominal figure for a lot at the time of its analysis, commonly obtained by the spectrophotometric or conductometric methods, which read higher than HPLC because they include oxidation products,19,20 and it says nothing about the history of the particular package. The S1 labels therefore cannot validate the absolute scale, and they are consistent with it only in the weaker sense that just three S1 samples exceed their declared values, Nelson Sauvin (008) by 6 %, Sterling (006) by 20 % and Belma (025) by 43 %.

The exception is sample 025 (Belma, declared 8.9 %), for which 12.75 % was found, 1.43 times the label, whereas the other Belma, 033, gave 6.77 %. The two carry the same label and the same recorded mass, and their compositions differ: the α/β ratio is 3.56 in 025 and 2.41 in 033, and the cohumulone fraction 23.3 against 26.2 %. They are therefore not two spectra of one preparation, and either the label, the recorded mass or the bottle of 025 is in error; the sample should be re-weighed and re-run before its result is used.

A4.4 Results

The contents of the six homologs, the α- and β-acid totals, the ratios and the terpene contents of all 49 samples are given in Table A6, and the homolog contents are shown in Figure A6. The α-acid contents range from 0.82 % in the Savinjski Golding (023) to 25.96 % in the Cryo Simcoe (041), with a median of 8.23 %, and the β-acid contents from 1.09 % (023) to 13.48 % in the Cryo Cascade (040), with a median of 3.99 %. The molar α/β ratio spans more than a factor of ten, from 0.43 in the Hersbrucker (028) to 4.98 in the Bravo (016). The cohumulone fraction of the α-acids runs from 17.0 % in the Cryo Simcoe (18.0 % by line fitting) to 50.0 % in Vic Secret (57.4 % by line fitting, the fixed windows understating high cohumulone fractions for the reason given in Section A3.2), and the colupulone fraction of the β-acids from 37.0 % in Crystal to 74.7 % in Vic Secret. The myrcene-type diene ranges from 0.10 % to 2.83 % (Cryo Mosaic, 042) and α-humulene from indistinguishable from zero to 1.68 % (042).

Table A6. Direct qNMR results for the 49 samples (% w/w, as-is basis): α-acid homologs, β-acid homologs (major + minor tautomers), totals, molar α/β ratio, co-homolog fractions (mol %), myrcene-type diene (as myrcene), α-humulene, and the ratio of measured to declared α-acids.
SampleCo-hum.Hum.Ad-hum.ΣαCo-lup.Lup.Ad-lup.Σβα/βco-α %co-β %Myrc.Hum-eneα meas./decl.
001 Magnum2.296.071.539.881.982.580.625.182.1723.939.00.550.670.80
002 Galena3.145.021.509.673.381.770.615.761.9133.459.50.430.410.83
003 Vic Secret4.613.181.639.423.560.800.454.812.2350.074.70.160.450.57
004 Idaho 72.874.091.278.231.420.720.282.423.8735.859.60.330.300.61
005 Ariana2.453.030.866.342.461.320.314.091.7639.661.00.160.220.61
006 Sterling1.605.131.077.812.342.480.725.541.6021.243.11.010.631.20
007 Summit2.495.461.389.331.911.630.443.992.6627.448.80.440.220.67
008 Nelson Sauvin2.786.741.5811.102.702.290.645.642.2425.848.80.530.491.06
009 German (unsp.)0.371.060.271.691.401.520.543.470.5622.341.30.190.270.48
010 Waimea2.727.721.7112.142.081.960.524.553.0323.146.50.760.240.84
011 Azacca2.865.901.6110.371.621.300.383.303.5728.450.00.530.350.76
012 Crystal0.391.030.231.651.131.600.383.120.6024.237.00.260.160.75
013 Trad. German0.921.920.593.431.070.910.342.321.6927.447.00.190.120.67
014 Enigma5.225.642.1112.972.781.100.474.343.4041.264.70.450.380.78
015 Sultana2.364.771.498.611.160.910.312.384.1128.149.60.670.430.69
016 Bravo3.285.311.8710.461.260.810.312.394.9832.253.70.610.310.74
017 N. Brewer1.381.920.633.931.540.720.272.541.7636.161.60.17< 0.10.68
018 Centennial1.543.930.956.420.950.830.272.053.5524.647.30.520.210.64
019 Saaz0.401.070.331.810.981.120.412.510.8222.940.00.13< 0.10.65
020 Chinook1.873.940.816.611.090.930.252.273.3129.049.00.300.230.50
021 Spalt0.390.920.261.571.101.060.352.510.7125.544.50.16< 0.10.75
022 Amarillo1.605.250.787.631.942.630.465.031.7321.639.40.550.390.83
023 Sav. Golding0.270.440.100.820.560.380.151.090.8534.252.50.100.150.31
024 Hallertau0.441.270.322.031.051.190.382.610.8822.340.80.180.200.73
025 Belma2.887.872.0012.751.911.620.554.073.5623.347.80.350.451.43
026 Magnum1.844.141.027.001.551.230.383.162.5227.049.80.350.200.79
027 Cluster1.852.860.885.592.151.150.413.711.7133.958.70.25< 0.10.76
028 Hersbrucker0.200.630.150.980.981.210.412.600.4321.238.40.200.180.58
029 Perle1.131.890.633.661.080.670.302.062.0231.753.50.240.310.72
030 Hall. Blanc1.604.260.926.781.561.820.463.832.0224.341.50.25< 0.10.89
031 Herkules3.886.981.8812.751.601.240.363.194.5531.350.90.590.500.77
032 Cascade1.122.040.483.641.751.500.373.621.1531.549.20.34< 0.10.65
033 Belma1.724.070.976.771.551.210.433.192.4126.249.40.240.290.76
034 Hersbrucker0.260.840.221.311.111.290.422.820.5320.340.20.140.210.77
035 Nugget3.188.581.9613.711.541.290.453.284.7523.947.80.720.490.90
036 Huell Melon0.831.850.493.182.542.210.615.360.6726.948.30.240.100.74
038 Centennial3.9311.422.9818.332.842.520.786.133.3922.147.12.511.240.99
039 Amarillo3.8712.962.9019.735.236.281.3412.851.7520.341.52.041.671.23
040 Cascade3.106.431.6311.156.625.521.3413.480.9428.650.01.810.990.94
041 Simcoe4.2718.213.4825.962.843.440.887.164.1217.040.51.381.401.01
042 Mosaic4.9015.595.1225.622.992.600.956.544.4419.846.62.831.681.19
043 Wakatu2.203.901.117.204.832.711.048.570.9531.357.20.930.451.04
044 Nelson Sauvin2.657.561.5311.742.642.460.645.742.3223.346.80.820.771.03
045 Wakatu2.524.611.328.445.172.941.059.151.0530.657.31.070.521.10
046 HBC 10191.765.481.258.482.652.380.925.951.6221.445.50.660.600.88
047 Waimea4.309.852.5016.653.452.640.786.872.7526.651.11.400.811.16
048 Nectaron4.069.322.4515.822.922.520.706.142.9326.448.41.191.151.16
049 Wakatu1.362.320.694.373.882.030.796.700.7432.058.80.470.560.82
050 HBC 10192.456.261.6010.313.902.811.097.791.5024.550.90.940.740.91
Provisional absolute scale (Section A5): contents carry an estimated uncertainty of about ±10 %; ratios and homolog fractions do not depend on the calibration. Line fitting indicates that the fixed windows overstate the adhumulone fraction of the Cryo pellets (038–042) by 3.7–5.0 points and understate their cohumulone fraction by 1.0–4.8 points, and understate the cohumulone fraction of the high-cohumulone pellets 003, 005 and 014 by 4.4–7.4 points (Section A3.2). Hum-ene: α-humulene; values below 0.1 % are not distinguishable from zero. Sample 025: see Section A4.3.
Contents of the three α-acid homologs (upper) and of the three β-acid homologs, major plus minor tautomers (lower), in the 49 hop samples by direct qNMR (% w/w, as is), ordered by total α-acids; bars…
Figure A6. Contents of the three α-acid homologs (upper) and of the three β-acid homologs, major plus minor tautomers (lower), in the 49 hop samples by direct qNMR (% w/w, as is), ordered by total α-acids; bars mark the declared α-acid content.

The Cryo pellets contained 2.6–3.1 times the α-acids, 2.6–3.7 times the β-acids and 3.7–5.4 times the myrcene-type diene of the retail pellets of the same cultivars in S1, whereas their declared α-acid contents were only 1.75–2.12 times higher. The enrichment of about twofold that the process is designed to give5 accounts for the declared ratios, and the larger measured ratios reflect the lower contents of the older S1 pellets, in the essential oil above all, rather than any excess in the Cryo products, whose ratio to the declared value is 1.07 on average. Their composition, on the other hand, is that of the parent cultivar: the α/β ratios of the Cryo and retail pellets are 3.39 and 3.55 for Centennial, 1.75 and 1.73 for Amarillo and 0.94 and 1.15 for Cascade.

A4.5 The ICE-4 Extract as an Absolute Check

The ICE-4 extract should have provided the one absolute validation of the procedure, and it did not (Table A3). With the recorded mass of 95.4 mg, bottle B gives 22.85 % α-acids and 15.23 % β-acids, 54 % and 57 % of the certified values, and bottle A 10.38 % and 6.92 %, 24 % and 26 %. Neither agrees. Degradation of the reference material cannot be excluded, but it is not suggested by the spectrum: the homolog composition matches the certificate to within one percentage point, the molar α/β ratio is within 7 % of it, and the oxidation-sensitive cluster W10 is no larger relative to the α-acids (0.049) than in the hops. The two recording irregularities noted in Section A2.4 are the more probable explanations: the ICE-4 spectrum is assigned to neither bottle, and its printed mass is that of sample 026. A mass of 51–55 mg with bottle B would reproduce both certified values within about 7 %, but that is conjecture. The result has to be recorded as a failed check, and the repetition of the ICE-4 measurement with a documented mass and a DMS calibrant added directly to the tube is the most valuable single experiment that remains to be done.

A4.6 Same-Cultivar Comparisons

Ten cultivars are represented more than once, and their comparisons, given in Table A7, separate what is fixed by the genotype from what is not. The clearest case is Wakatu, of which three harvests were examined. The α-acid content was 4.37 % in the 2022 sample, 7.20 % in 2023 and 8.44 % in 2024, in the same order as the declared values (5.3, 6.9 and 7.7 %), and the α/β ratio rose from 0.74 to 1.05, whereas the cohumulone fraction (32.0, 31.3 and 30.6 %) and the colupulone fraction (58.8, 57.2 and 57.3 %) varied by little more than a percentage point. The homolog distributions of a cultivar are thus reproduced across harvests within the uncertainty of the windows, as Molyneux and Wong concluded from eight Cascade lots in 1975,4 while the amounts and the α/β ratio depend on the lot and its age; the lower α/β ratio of the oldest Wakatu may reflect lot-to-lot variation or a preferential loss of α-acids on storage, and three samples cannot distinguish the two. The two HBC 1019 samples (2022 and 2024) differ by 3.1 points in cohumulone and by 5.4 points in colupulone (45.5 against 50.9 %), and the retail and 2023 Waimea samples by 3.5 and 4.6 points, differences several times larger than those among the three Wakatu harvests whose cause a single pair of samples cannot establish. The retail and 2023 Nelson Sauvin samples are nearly indistinguishable (α-acids 11.10 and 11.74 %, α/β 2.24 and 2.32), as are the two Hersbrucker samples in composition.

Two pairs do not behave as replicate samples of one cultivar should. The Belma pair has been discussed above. The Magnum pair differs by 10.7 points in its colupulone fraction (39.0 % in the German Magnum, 001, and 49.8 % in 026), a difference far outside the within-cultivar range of the other pairs; the composition of 026 is, on the other hand, almost identical to that of the Belma 033 (cohumulone 27.0 and 26.2 %, colupulone 49.8 and 49.4 %, α/β 2.52 and 2.41), which strengthens the suspicion expressed in v1.1 that its label, which carries the Belma value of 8.9 %, was transcribed from the neighboring sample. Both questions should be resolved from the laboratory records before the samples are used as references.

Table A7. Cultivars represented by more than one sample: declared and measured α-acids, β-acids, α/β, co-homolog fractions and myrcene-type diene, and the distance between the two samples in the first three principal components of the compositional data set (as a fraction of the median distance between all pairs) with the rank of the partner among the nearest neighbors.
CultivarSamplesFormsDeclared α (%)α (%)β (%)α/βco-α (%)co-β (%)Myrcene (%)Distance / medianRank
Magnum001 / 026Pellet / Pellet12.3 / 8.99.88 / 7.005.18 / 3.162.17 / 2.5223.9 / 27.039.0 / 49.80.55 / 0.350.314
Nelson Sauvin008 / 044Pellet / Pellet10.5 / 11.411.10 / 11.745.64 / 5.742.24 / 2.3225.8 / 23.348.8 / 46.80.53 / 0.820.359
Waimea010 / 047Pellet / Pellet14.5 / 14.312.14 / 16.654.55 / 6.873.03 / 2.7523.1 / 26.646.5 / 51.10.76 / 1.400.4213
Centennial018 / 038Pellet / Cryo10.1 / 18.56.42 / 18.332.05 / 6.133.55 / 3.3924.6 / 22.147.3 / 47.10.52 / 2.510.6920
Amarillo022 / 039Pellet / Cryo9.2 / 16.17.63 / 19.735.03 / 12.851.73 / 1.7521.6 / 20.339.4 / 41.50.55 / 2.040.363
Belma025 / 033Pellet / Pellet8.9 / 8.912.75 / 6.774.07 / 3.193.56 / 2.4123.3 / 26.247.8 / 49.40.35 / 0.240.357
Hersbrucker028 / 034Pellet / Pellet1.7 / 1.70.98 / 1.312.60 / 2.820.43 / 0.5321.2 / 20.338.4 / 40.20.20 / 0.140.301
Cascade032 / 040Pellet / Cryo5.6 / 11.93.64 / 11.153.62 / 13.481.15 / 0.9431.5 / 28.649.2 / 50.00.34 / 1.810.7818
Wakatu043 / 045Pellet / Pellet6.9 / 7.77.20 / 8.448.57 / 9.150.95 / 1.0531.3 / 30.657.2 / 57.30.93 / 1.070.121
Wakatu043 / 049Pellet / Pellet6.9 / 5.37.20 / 4.378.57 / 6.700.95 / 0.7431.3 / 32.057.2 / 58.80.93 / 0.470.464
Wakatu045 / 049Pellet / Pellet7.7 / 5.38.44 / 4.379.15 / 6.701.05 / 0.7430.6 / 32.057.3 / 58.81.07 / 0.470.587
HBC 1019046 / 050Pellet / Pellet9.6 / 11.38.48 / 10.315.95 / 7.791.62 / 1.5021.4 / 24.545.5 / 50.90.66 / 0.940.191
Wakatu: 043 (2023), 045 (2024), 049 (2022). HBC 1019: 046 (2022), 050 (2024). Nelson Sauvin 044 and Waimea 047: 2023. Rank 1 = the partner is the nearest neighbor among the other 48 samples.

A4.7 The Co-Homolog Relationship

The linear relationship between the colupulone fraction of the β-acids and the cohumulone fraction of the α-acids reported in v1.1 for the 36 survey samples (slope 1.33, intercept 11.6 percentage points, r² = 0.90) is reproduced exactly by the S1 samples of the present data set (slope 1.327, intercept 11.56, r² = 0.897), which confirms that the major windows used here are those of the v1.1 archive. With all 49 samples the slope is 1.23, the intercept 15.4 and r² 0.84 (Figure A7). Part of the reduction is caused by the Cryo pellets, whose cohumulone fractions are understated by the fixed windows (Section A3.2); with their line-fit values the slope becomes 1.27 and r² 0.86. The three Wakatu samples lie three to four points above the line, with colupulone fractions of 57–59 % at cohumulone fractions of 31–32 %, and the German Magnum (001), the retail Cascade (032) and Crystal (012) lie five to nine points below it; the relationship is therefore a useful consistency check but not a tight one across genetically diverse cultivars.

Colupulone fraction of the β-acids (major tautomer signals) against the cohumulone fraction of the α-acids for the 49 samples, with the regression line of this data set and that of v1.1
Figure A7. Colupulone fraction of the β-acids (major tautomer signals) against the cohumulone fraction of the α-acids for the 49 samples, with the regression line of this data set and that of v1.1. Heavy outlines mark set S2; arrows show the shift of the Cryo pellets when their cohumulone fractions are taken from line fitting instead of the fixed windows.

A4.8 Terpenes

The myrcene-type diene and α-humulene contents follow the product and its age more closely than the cultivar. The median myrcene-type content is 0.31 % in S1, 0.94 % in the fresh S2 pellets and 2.04 % in the Cryo pellets, and α-humulene 0.24, 0.67 and 1.40 % respectively. Both are hydrocarbons that are lost from hops by evaporation and oxidation during storage, and both can be lost from the sample during extraction, so that the NMR values are lower bounds to the contents of the hop as sampled. The humulene values carry the additional uncertainty described in Section A3.6. Within these limits the ratio of the myrcene-type signal to the α-acids, which rises from 0.15 in S1 to 0.23 in the S2 pellets and 0.29 in the Cryo pellets (medians), is a useful index of freshness, and it is this ratio, rather than any geographic property, that drives the oil-related dimension of the chemometric models of Section A8.

References Cited in This Section

Numbered as in the full list on the References page.

4.
Molyneux, R. J.; Wong, Y.-I. Nuclear magnetic resonance spectroscopic determination of α- and β-acid homolog composition in hops. J. Agric. Food Chem. 1975, 23 (6), 1201–1204. https://doi.org/10.1021/jf60202a012
5.
Yakima Chief Hops. Commercial Cryo Hops®; product information; Yakima, WA. https://www.yakimachief.com/commercial-cryo-hops (accessed September 23, 2026).
10.
Pauli, G. F.; Gödecke, T.; Jaki, B. U.; Lankin, D. C. Quantitative ¹H NMR. Development and potential of an analytical method: an update. J. Nat. Prod. 2012, 75 (4), 834–851. https://doi.org/10.1021/np200993k
11.
Malz, F.; Jancke, H. Validation of quantitative NMR. J. Pharm. Biomed. Anal. 2005, 38 (5), 813–823. https://doi.org/10.1016/j.jpba.2005.01.043
12.
Mo, H.; Raftery, D. Solvent Signal as an NMR Concentration Reference. Anal. Chem. 2008, 80 (24), 9835–9839. https://doi.org/10.1021/ac801938j
13.
Pierens, G. K.; Carroll, A. R.; Davis, R. A.; Palframan, M. E.; Quinn, R. J. Determination of analyte concentration using the residual solvent resonance in ¹H NMR spectroscopy. J. Nat. Prod. 2008, 71 (5), 810–813. https://doi.org/10.1021/np8000046
14.
Wider, G.; Dreier, L. Measuring Protein Concentrations by NMR Spectroscopy. J. Am. Chem. Soc. 2006, 128 (8), 2571–2576. https://doi.org/10.1021/ja055336t
15.
Akoka, S.; Barantin, L.; Trierweiler, M. Concentration Measurement by Proton NMR Using the ERETIC Method. Anal. Chem. 1999, 71 (13), 2554–2557. https://doi.org/10.1021/ac981422i
16.
Fulmer, G. R.; Miller, A. J. M.; Sherden, N. H.; Gottlieb, H. E.; Nudelman, A.; Stoltz, B. M.; Bercaw, J. E.; Goldberg, K. I. NMR Chemical Shifts of Trace Impurities: Common Laboratory Solvents, Organics, and Gases in Deuterated Solvents Relevant to the Organometallic Chemist. Organometallics 2010, 29 (9), 2176–2179. https://doi.org/10.1021/om100106e
17.
Nickerson, G. B.; Likens, S. T. Hop Storage Index. J. Am. Soc. Brew. Chem. 1979, 37 (4), 184–187. https://doi.org/10.1094/asbcj-37-0184
18.
Mikyška, A.; Krofta, K. Assessment of changes in hop resins and polyphenols during long-term storage. J. Inst. Brew. 2012, 118 (3), 269–279. https://doi.org/10.1002/jib.40
19.
American Society of Brewing Chemists. Hops-6: α-Acids and β-acids in hops and hop pellets by spectrophotometry. In ASBC Methods of Analysis; ASBC: St. Paul, MN (online edition).
20.
American Society of Brewing Chemists. Hops-14: α-Acids and β-acids in hops and hop extracts by HPLC (International Method). In ASBC Methods of Analysis; ASBC: St. Paul, MN (online edition).
Back to top