Abstract and Document Details
Document details and provenance
| Document | Technical white paper PNA-WP-2026-SPIRITS-01, version 1.5 |
|---|---|
| Author and owner | John C. Edwards, Ph.D., Process NMR Associates, LLC, 84 Patrick Lane, Ste 115, Poughkeepsie, NY 12603 |
| Generated | 23 October 2026 |
| Revision | Version 1.5 (23 October 2026) adds Section 2.4, in which a library of 71 reference spectra of 45 pure congeners and an eleven-step mixture series, decoded from a further MestReNova document, are shown in stacked plots, the solvents their solubility required (D₂O, aqueous ethanol, D₂O–CD₃OD, CDCl₃, DMSO-d₆) and the resulting shift differences are discussed, and the references are aligned with the intact spirits (four figures, two tables). The whisky-lactone screen of Section 6.4 was rerun at the positions measured in the reference recorded as a spirit, which lie 0.16 ppm from the CDCl₃ positions searched in version 1.4; the limits of detection, the figure and the text were revised, and the earlier statement that the cis methyl doublet offered a limit near 7 mg/L was withdrawn, because that doublet falls on the ethanol ¹³C satellite lines. Section 10.3 adds a second round of PLS calibrations after removal of flagged and statistically outlying calibration products, with two figures and a table, and Section 10.4 examines the choice of the number of latent variables (formal criteria applied to repeated cross-validations, the content of every component and models restricted to the resonances of each analyte), with three figures and a table; the original calibrations of Section 10.2 are unchanged and Sections 10.3–10.7 became 10.5–10.9. Section 3.1 states the accounting from 204 spectra to 166 products and 163 chemometric products explicitly. Seventeen references were added, each checked against its CrossRef record. Version 1.4 (23 October 2026) corrects the flip angles of the acquisition: the 90° pulse width of the probe is 13 µs, so that the 7, 8 and 9 µs pulses correspond to 48°, 55° and 62° rather than to the 63–81° inferred from the pw90 parameter stored in the procpar files, and the relaxation estimates of Sections 5.4 and 11.2 were recomputed accordingly with the T₁ of maleic acid in D₂O, 6.1 s; no relaxation correction had been applied to any result. All quantitative results, and the statistics, figures and tables derived from them, were recalculated and the text was updated to the new values. The observability of the oak lactones is examined in Section 6.4, with a new figure, together with an oak-associated doublet at 4.47/4.50 ppm that was added to the assignment table, and seven references were added. Version 1.3 (23 October 2026) placed the Process NMR Associates logo on the title page and in the running headers and extended the reference list from 60 to 97 entries: 14 references verified for version 1.0 but not then cited and 23 new ones, each checked against its CrossRef record or the official regulatory text, are cited where they support statements of the report (spirit production and congeners, the chemometric methods, outlier statistics and class models, caramel and furanic aldehydes, fusel oil, surrogate alcohol, propylene glycol and coffee constituents). A few sentences were extended to carry them and the citation numbers changed throughout; no result was changed. Version 1.2 (22 October 2026) added Section 12, a study of counterfeit spirits and undeclared additives: eleven further spectra (four additized vodkas, six methanol–ethanol–water mixtures and a caramel adjunct) decoded from a second MestReNova document; quantification of the mixtures by internal normalization and against the maleic acid found in them; projection of the counterfeit samples into the models of Sections 9 and 10 with Hotelling T² and Q limits, class models, Ward clustering and joint UMAP and t-SNE embeddings; and a propylene glycol screen of all 204 spectra. The abstract, the introduction, the table of recommendations, the conclusions (now Section 13) and the summary were extended, and the 3-D score viewer shows the counterfeit samples. The results of Sections 3–11 are unchanged; one reference was withdrawn (Barker and Rayens, 2003) and one added (Botelho et al., 2020), which renumbers the citations from 51 onward. Version 1.1 superseded version 1.0 of the same date. Methanol was re-quantified with a two-line model (methanol singlet with the upfield line of the isobutanol CH₂O doublet; Sections 3.5, 4.5 and 6.2); limits of quantification were re-estimated through the fit models (Sections 3.5 and 6.1); the clipped methanol peaks of two set B traces were reconstructed and the binned blocks rebuilt with adaptive bins (Section 3.7); all chemometric and regression results were recomputed (Sections 9 and 10), and the methanol calibration, which had failed, was repaired (Section 10.2). The values of version 1.0 are listed beside the revised values in the companion workbook (sheets Methanol_Revision and LOQ_Revision). |
| File name | PNA-WP-2026-SPIRITS-01_Intact_Spirits_qNMR_v1.5.docx |
| Companion files | PNA-WP-2026-SPIRITS-01_Companion_Workbook.xlsx (all inputs, calculations and outputs, including every chemometric and regression data set, with live qNMR formulas); PNA-WP-2026-SPIRITS-01_Congener_Reference_Library.xlsx (Section 2.4); PNA-WP-2026-SPIRITS-01_PLS_Outlier_Round2.xlsx (Sections 10.3 and 10.4); Oak_Lactone_Screen.xlsx (Section 6.4); PNA-WP-2026-SPIRITS-01_Code_and_Data.zip (Python and Node.js source code, processed data blocks, reference records); interactive 3-D score viewer for PCA, UMAP and t-SNE (published page, also supplied as PNA-WP-2026-SPIRITS-01_3D_Score_Viewer.html) |
| Spectral data | 204 ¹H NMR spectra of intact spirits acquired by Process NMR Associates on the Varian Mercury-MVX 300 (JCE-PNA-MVX300): 106 decoded from the MestReNova project supplied for the 106-sample set and 98 vectorized from the PDF plots supplied for the 98-sample set (ethanol on scale, congener expansion and higher-alcohol methyl window); a further 11 spectra of counterfeit samples and of a caramel adjunct (Section 12) decoded from a second MestReNova document; and 71 reference spectra of pure congeners and an eleven-step mixture series (Section 2.4) decoded from a third |
| AI model identity | claude-opus-5-5 (configured model identifier as reported by the session; the serving model may differ) |
| Tools engaged | Python 3.13 with NumPy, SciPy, pandas, scikit-learn, umap-learn, matplotlib, PyMuPDF, glymur/OpenJPEG and imagecodecs; Node.js with docx 9.8.1; LibreOffice (rendering and table-of-contents pagination); CrossRef REST API and the official publisher or regulator pages (reference verification); plotly.js 2.35.3 (3-D viewer) |
| AI-assistance statement | The decoding of the spectral files, the quantitative calculations, the chemometric and regression computations, the figures and tables and the draft text were produced with AI assistance from the author's spectra, product descriptions and stated preparation and acquisition parameters. Sample preparation, NMR acquisition, the product descriptions (including label alcohol contents) and the descriptions and recipes of the counterfeit samples are the author's. Statements identified in the text as hypotheses, notably the proposed causes of the campaign-dependent ethanol deviation, have not been tested experimentally and require the author's confirmation before they are relied upon. |
| Reference verification | Every cited journal article, book and proceedings chapter was checked against its CrossRef record (title, full author list, journal or book, volume, pages and DOI); regulations and standards were checked against the official texts. Only verified references are cited (121 in version 1.5). |
| Numerical verification | All numbers quoted in the text are generated from the result files by the build scripts; the qNMR equations were re-implemented as live formulas in the companion workbook and recalculated without error against the Python results. |
Abstract
A total of 204 ¹H NMR spectra of intact brown spirits, each prepared from 175 µL of spirit, 450 µL of D₂O and 50 µL of a maleic acid stock (4.75 mg of internal standard) and acquired at 300 MHz with 7–9 µs pulses (48–62° for the 13 µs 90° pulse of the probe, against a nominal 45°) and, for most spectra, a 15 s relaxation delay, were recovered from a MestReNova project (106 spectra decoded losslessly from the native document) and from vector PDF plots (98 spectra) and were merged into a single data set of 166 products: bourbons, American rye whiskeys, Scotch, Irish, Taiwanese and other whiskies, rums, cognacs and other brandies. Ethanol was quantified from its methyl triplet against the internal standard and expressed as % ABV and % w/w through the OIML R 22 density function. Methanol was fitted together with the upfield line of the isobutanol CH₂O doublet, which lies only 0.8–1.9 Hz from it and moves closer as the alcoholic strength rises; acetic acid, ethyl acetate, succinic acid, 2-phenylethanol, formic acid, furfural, 5-hydroxymethylfurfural, acetaldehyde, glucose and sucrose were quantified by constrained lineshape fitting, with limits of quantification propagated from the noise of each spectrum through its fit model; and the four principal higher alcohols were resolved by deconvolution of the 0.825–0.915 ppm methyl envelope with globally refined chemical shifts and couplings. Chemical-shift and coupling tables, annotated spectra and statistical total correlation spectroscopy were used to support the assignments; those of formic acid, furfural and 5-hydroxymethylfurfural rest on a single signal and remain tentative. A library of 71 reference spectra of 45 pure congeners, recorded in D₂O, aqueous ethanol, D₂O–CD₃OD, CDCl₃ or DMSO-d₆ as their solubility required, and an eleven-step mixture series were aligned with the intact spirits: solvent shifts of −0.15 to +0.22 ppm make non-aqueous references unsuitable for placing signals in a spirit, aqueous references lie a median 0.018 ppm downfield of the spirit positions, and references prepared as a spirit place them within about 0.005 ppm. They supported a tentative gallic acid singlet at 7.150 ppm in 26 of 101 products and placed the whisky-lactone signals 0.16 ppm from their CDCl₃ positions, where the lactones remain below the reach of the measurement.
For the 125 labeled products with a valid internal-standard dose, the qNMR value was a median 0.8 % (relative) below the label, but the deviation proved to depend on the acquisition campaign, with medians from −16.4 % to +1.5 % for the months with three or more labeled products; the pattern points to the internal-standard dosing chain rather than to a constant physical effect, and congener concentrations expressed per hectoliter of absolute alcohol are immune to it. The maleic acid internal standard was found to esterify in the tube to monoethyl maleate (median 0.5 mol %, maximum 5.0 mol %), fixed integration windows were shown to cross-contaminate the acetic acid and ethyl acetate singlets and to over-read methanol, and a single-line methanol model was found to overestimate methanol two-fold or more in 52 of the 163 products of the chemometric set; the two-line model, checked against the isobutanol methyl doublet (r = 0.986), lowered the median methanol content from 73 to 49 mg/L. Succinic acid, 2-phenylethanol, the aldehydes and sucrose lie below their limits of quantification in most products. Four data blocks (regional integrals, adaptively binned spectra, higher-alcohol fit areas and congener concentrations) were submitted to a pre-processing screen and to PCA, UMAP, t-SNE, Ward clustering, k-nearest neighbors, support vector machines and PLS-DA under nested cross-validation. Seven-class balanced accuracies of 0.63–0.71 were obtained with radial-basis support vector machines (chance 0.14; the PLS-DA models, which alone were permutation-tested, gave p ≤ 0.016), the fitted higher-alcohol areas and the congener concentrations performing as well as the binned spectra, whose results were shown to be strongly confounded with the data route; cognac and Taiwanese single malt were recognized with recalls of 0.80–1.00, whereas bourbon and American rye whiskey were extensively confused. The binned spectra used bin edges placed in spectral valleys, because an equidistant 0.01 ppm grid cut 12 of 34 assigned lines and shifting it by half a bin cut 16, including both acetyl singlets. PLS and support-vector regression of the binned spectra against the qNMR concentrations, validated on 20 withheld products, reproduced methanol, acetic acid, ethyl acetate and the higher alcohols with R²P = 0.93–0.997 for PLS; a second round after removal of flagged and outlying calibration products lowered the validation errors significantly for acetic acid and isobutanol, most of the gain being traceable to one product prepared with half the internal-standard dose; formal tests on repeated cross-validations confirmed the 7–17 latent variables of the cleaned models for most targets, but the components beyond about the tenth were shown to draw mainly on the signals of other congeners, and models restricted to each analyte's resonances matched or improved the predictions with fewer components for succinic acid, acetic acid, 1-propanol and the total higher alcohols; the methanol calibration, which had failed in the first version of this analysis, worked only after a clipped trace had been repaired and the reference values corrected, and moving the bin edges off the methanol line reduced its error by a further third, whereas succinic acid, mostly below its limit of quantification, was predicted from correlated congeners rather than measured. Because the reference values were derived from the same spectra, these calibrations establish internal consistency rather than accuracy against an independent method. A gravimetric internal-standard protocol with campaign check standards, an inert internal standard or timed acquisition, lineshape-fit quantification with every overlapping line modeled in place of fixed windows, native-format data for every spectrum and GC-FID cross-validation are recommended.
A supplementary set of counterfeit samples was examined with the same procedures and models. Four vodkas to which a commercial bourbon, Kentucky bourbon or dark rum flavor or 0.1 % of caramel had been added carried 1.0–40 g/hL a.a. of higher alcohols, against at least 292 g/hL a.a. in every authentic bourbon, yet they lay inside or only marginally outside the 99 % limits of the global principal component models, whose false-alarm rates on the authentic products were 3.7–9.8 %, except where the propylene glycol of one of them reached the binned spectra; the seven-class support vector machine called all four rum. In these samples a class model of authentic bourbon rejected all four, although it also accepted 62 of the 146 other authentic products, and any threshold on the higher alcohols between 40 and 220 g/hL a.a. separated them from all 79 bourbons, ryes, Scotch, Taiwanese and Japanese whiskies of the set; neither test exposed the vodka flavored as rum, nor can either be expected to, because the authentic rums extend to near-neutral column spirits, and the threshold would not survive the addition of fusel oil. Methanol in mixtures with ethanol and water was quantified by internal normalization of the ¹H integrals and, because the tubes proved to contain the maleic acid standard, by Equation 1 as well, which agreed with the recipes to within about 10 % (relative); the methanol fraction of the alcohol, which needs neither the standard nor the water signal, agreed to within 1.4 percentage points. A screen of all spectra found propylene glycol, the carrier solvent of liquid flavorings, in an orange-flavored whiskey (13.7 g/L), a Texas blended whiskey (2.8 g/L) and, probably, a sweetened rum whose higher alcohols are those of a neutral spirit.